http://hdl.handle.net/2289/3213| Title: | Imidazole modified C-paste electrode in the determination of trace amounts of Mercury |
| Authors: | Kannan, R. Rajagopalan, S.R. Lakshminarayanan, V. |
| Issue Date: | 1995 |
| Publisher: | The Electrochemical Society of India |
| Citation: | Journal of the Electrochemical Society of India, 1995, Vol.44, p90-96 |
| Abstract: | The imidazole-modified C-paste electrode was shown lo accumulate mercury (II) ions non- electrochemically from nitrate containing solutions at pH 7.0-7.3. The accumulated mercury gave well defined voltammograms, when subjected to voltammetric scale in a medium of potassium thiocyanate and perchloric acid; these are suitable for determination of mercury. In comparison with pure C-paste electrodes, the imidazole-CME is shown to offer improved resistance to interference by diverse types of ions present in natural waters. The method offered a sensitivity of 1.75 μA/ppm and a detection limit of 0.1 ppm (lor 15 min. accumulation time). The relative standard deviation was 12%. |
| Description: | Restricted Access. |
| URI: | http://hdl.handle.net/2289/3213 |
| ISSN: | 0013-466X |
| Copyright: | 1995 The Electrochemical Society of India |